This is an open reference for anyone connecting lab and point-of-care analysers to a LIS, EHR or practice system: which protocol each analyser speaks, how it physically connects, and whether Catenix has a driver or a connection guide for it. It covers 135 analysers from 47 manufacturers and 25 clinical record systems. Use it freely, with credit to Catenix (CC BY 4.0).
Download: JSON · CSV · last updated 2026-11-18.
Message formats vary by software version, so the exact format is confirmed on each unit at setup. Analyser not listed? Any analyser that sends HL7 v2, ASTM E1394 or POCT1-A2 can be connected, and Catenix builds the driver for anything else, usually within a couple of days.
Analysers
Blood gas
| Analyser | Protocol | Physical link | Catenix |
|---|---|---|---|
| Abbott i-STAT | HL7 v2 | through the manufacturer's data-management software, dock or hub; over the site network (TCP/IP) | Driver built |
| Abbott i-STAT Alinity | POCT1-A2 | over the site network (TCP/IP) | Driver built |
| Edan i15 Blood Gas and Chemistry Analyser | Confirmed at setup | through the manufacturer's data-management software, dock or hub | Connection guide (standard protocol) |
| EXIAS e|1 | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Nova Prime Plus | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Nova Stat Profile Prime Plus | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Radiometer ABL800 FLEX | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Radiometer ABL90 FLEX | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Roche cobas b 123 POC system | ASTM E1381/E1394; POCT1-A2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Roche cobas b 221 | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Siemens RAPIDLab 1200 | Confirmed at setup | RS-232 serial | Connection guide (standard protocol) |
| Siemens RAPIDPoint 500 | Confirmed at setup | through the manufacturer's data-management software, dock or hub; RS-232 serial, bridged to the network with a serial device server | Driver built |
| Siemens epoc | HL7 v2 | over the site network (TCP/IP) | Driver built |
| Werfen GEM Premier | HL7 v2 | over the site network (TCP/IP) | Driver built |
| Werfen GEM Premier 3500 | HL7 v2; ASTM E1381/E1394 | through the manufacturer's data-management software, dock or hub; RS-232 serial | Driver built |
| Werfen GEM Premier 5000 | HL7 v2; ASTM E1381/E1394 | through the manufacturer's data-management software, dock or hub; over the site network (TCP/IP) | Driver built |
Clinical chemistry
| Analyser | Protocol | Physical link | Catenix |
|---|---|---|---|
| Abbott ARCHITECT c8000 | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Abbott FreeStyle Precision Pro | Confirmed at setup | through the manufacturer's data-management software, dock or hub | Connection guide (standard protocol) |
| Abbott Piccolo Xpress | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Beckman Coulter AU480 | Confirmed at setup | RS-232 serial, bridged to the network with a serial device server | Connection guide (standard protocol) |
| Beckman Coulter AU5800 | Confirmed at setup | Confirmed at setup | Connection guide (standard protocol) |
| Beckman Coulter AU680 | Confirmed at setup | Confirmed at setup | Connection guide (standard protocol) |
| Beckman Coulter DxC 700 AU | Confirmed at setup | Confirmed at setup | Connection guide (standard protocol) |
| Diamond Diagnostics CareLyte | Confirmed at setup | Confirmed at setup | Driver built |
| Fujifilm DRI-CHEM NX500 | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Fujifilm DRI-CHEM NX600 | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| HORIBA Pentra C400 | ASTM E1381/E1394 | RS-232 serial, bridged to the network with a serial device server | Connection guide (standard protocol) |
| Medica EasyLyte | Confirmed at setup | RS-232 serial, bridged to the network with a serial device server | Driver built |
| Mindray BS-200 | Confirmed at setup | Confirmed at setup | Driver built |
| Mindray BS-240 | Confirmed at setup | Confirmed at setup | Driver built |
| Mindray BS-480 | Confirmed at setup | RS-232 serial | Driver built |
| Nova StatSensor Creatinine Meter | POCT1-A2 | through the manufacturer's data-management software, dock or hub; over the site network (TCP/IP) | Connection guide (standard protocol) |
| QuidelOrtho VITROS XT 7600 Integrated System | Confirmed at setup | RS-232 serial, bridged to the network with a serial device server | Connection guide (standard protocol) |
| Randox RX Daytona+ | ASTM E1381/E1394 | RS-232 serial, bridged to the network with a serial device server | Connection guide (standard protocol) |
| Roche Accu-Chek Inform II | POCT1-A2 | through the manufacturer's data-management software, dock or hub; over the site network (TCP/IP) | Connection guide (standard protocol) |
| Roche cobas c 111 | ASTM E1381/E1394 | RS-232 serial, bridged to the network with a serial device server | Connection guide (standard protocol) |
| Roche cobas c 311 | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Roche cobas pure integrated solutions | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Seamaty SD1 | HL7 v2 | over the site network (TCP/IP) | Driver built |
| Siemens Healthineers Dimension EXL 200 | Confirmed at setup | RS-232 serial | Connection guide (standard protocol) |
Coagulation
| Analyser | Protocol | Physical link | Catenix |
|---|---|---|---|
| Haemonetics TEG 6s | HL7 v2 | Confirmed at setup | Connection guide (standard protocol) |
| Roche CoaguChek Pro II | POCT1-A2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Roche CoaguChek XS Pro | POCT1-A2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Roche cobas t 511 | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Sysmex CA-660 | ASTM E1381/E1394 | RS-232 serial, bridged to the network with a serial device server | Connection guide (standard protocol) |
| Sysmex CS-1600 | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Werfen ACL Elite Pro | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Werfen ACL TOP 550 CTS | Confirmed at setup | Confirmed at setup | Connection guide (standard protocol) |
| Werfen GEM Hemochron 100 | Confirmed at setup | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Werfen Hemochron Signature Elite | POCT1-A2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
Glucose and HbA1c
| Analyser | Protocol | Physical link | Catenix |
|---|---|---|---|
| ACON On Call MultiPro | HL7 v2 | over the site network (TCP/IP) | Driver built |
| Abbott Afinion 2 | POCT1-A2 | over the site network (TCP/IP) | Driver built |
| Bio-Rad D-10 | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| EKF Diagnostics Quo-Test HbA1c Analyser | POCT1-A2 | through the manufacturer's data-management software, dock or hub; RS-232 serial | Connection guide (standard protocol) |
| Menarini HA-8180 | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Nova Allegro Analyser | HL7 v2; ASTM E1381/E1394; POCT1-A2 | RS-232 serial | Connection guide (standard protocol) |
| Nova StatStrip | HL7 v2 | through the manufacturer's data-management software, dock or hub; RS-232 serial | Connection guide (standard protocol) |
| Roche cobas b 101 system | POCT1-A2 | through the manufacturer's data-management software, dock or hub; over the site network (TCP/IP) | Connection guide (standard protocol) |
| Siemens DCA Vantage | ASTM E1381/E1394; POCT1-A2 | RS-232 serial | Driver built |
| Tosoh HLC-723 G8 | Manufacturer's own format | RS-232 serial | Driver built |
| Tosoh HLC-723G11 | Confirmed at setup | RS-232 serial | Driver built |
| Tosoh HLC-723GX | Confirmed at setup | RS-232 serial | Driver built |
Haematology
| Analyser | Protocol | Physical link | Catenix |
|---|---|---|---|
| Beckman Coulter DxH 500 | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Beckman Coulter DxH 900 | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Dymind DF55 | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Dymind DH-800 | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Dymind DH36 | Confirmed at setup | over the site network (TCP/IP) | Connection guide (standard protocol) |
| HORIBA ABX Micros ES 60 | HL7 v2; ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| HORIBA ABX Pentra 60 | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| HORIBA Yumizen H500 | HL7 v2 | RS-232 serial | Connection guide (standard protocol) |
| HemoCue Hb 201 DM | POCT1-A2 | through the manufacturer's data-management software, dock or hub; over the site network (TCP/IP) | Connection guide (standard protocol) |
| HemoCue WBC DIFF | POCT1-A2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Medonic M32 | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Mindray BC-20s | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Mindray BC-30s | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Mindray BC-5150 | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Mindray BC-5380 | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Mindray BC-5390 | Confirmed at setup | Confirmed at setup | Connection guide (standard protocol) |
| Nihon Kohden Celltac | Confirmed at setup | Confirmed at setup | Driver built |
| Ozelle EHBT-75 | HL7 v2 | Network (TCP/IP) | Driver built |
| Seamaty SMT-50 | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Sysmex KX-21N | Confirmed at setup | RS-232 serial, bridged to the network with a serial device server | Connection guide (standard protocol) |
| Sysmex XN-1000 | ASTM E1381/E1394 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Sysmex XN-550 | ASTM E1381/E1394; Manufacturer's own format | RS-232 serial | Connection guide (standard protocol) |
| Sysmex XP-300 | HL7 v2 | RS-232 serial | Connection guide (standard protocol) |
| Sysmex XQ-320 | HL7 v2; ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Sysmex pocH-100i | Confirmed at setup | RS-232 serial, bridged to the network with a serial device server | Connection guide (standard protocol) |
| Zybio Z3 | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
Immunoassay and cardiac
| Analyser | Protocol | Physical link | Catenix |
|---|---|---|---|
| Abbott ARCHITECT i2000SR | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Aidian QuikRead go | ASTM E1381/E1394; POCT1-A2 | RS-232 serial | Connection guide (standard protocol) |
| BD Veritor Plus System | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Beckman Coulter Access 2 | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Boditech AFIAS 6 | Confirmed at setup | over the site network (TCP/IP) | Driver built |
| DiaSorin LIAISON XL | Confirmed at setup | Confirmed at setup | Connection guide (standard protocol) |
| Getein 1600 Immunofluorescence Quantitative Analyser | Confirmed at setup | over the site network (TCP/IP) | Connection guide (standard protocol) |
| LumiraDx Instrument | HL7 v2 | over the site network (TCP/IP); Bluetooth | Connection guide (standard protocol) |
| PATHFAST Immunoanalyser | ASTM E1381/E1394 | RS-232 serial, bridged to the network with a serial device server | Connection guide (standard protocol) |
| Quidel Sofia 2 | ASTM E1381/E1394; POCT1-A2 | over the site network (TCP/IP) | Driver built |
| Quidel Triage MeterPro | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Radiometer AQT90 FLEX | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Roche cobas e 411 | ASTM E1381/E1394 | RS-232 serial | Driver built |
| Roche cobas h 232 POC system | POCT1-A2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Siemens Atellica VTLi | POCT1-A2 | through the manufacturer's data-management software, dock or hub; over the site network (TCP/IP) | Driver built |
| Siemens Healthineers ADVIA Centaur XPT | Confirmed at setup | RS-232 serial | Connection guide (standard protocol) |
| Siemens Healthineers IMMULITE 1000 | ASTM E1381/E1394 | Confirmed at setup | Connection guide (standard protocol) |
| Siemens Healthineers IMMULITE 2000 XPi | Confirmed at setup | Confirmed at setup | Connection guide (standard protocol) |
| Siemens Stratus CS 200 Acute Care Analyser | Confirmed at setup | RS-232 serial | Connection guide (standard protocol) |
| Snibe MAGLUMI 2000 Plus | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Snibe MAGLUMI 4000 Plus | Confirmed at setup | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Snibe MAGLUMI 800 | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Snibe MAGLUMI X3 | HL7 v2; ASTM E1381/E1394 | RS-232 serial | Driver built |
| Snibe MAGLUMI X8 | ASTM E1381/E1394 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Tosoh AIA-2000 | HL7 v2; ASTM E1381/E1394 | RS-232 serial | Driver built |
| Tosoh AIA-360 | ASTM E1381/E1394 | RS-232 serial | Driver built |
| Tosoh AIA-900 | ASTM E1381/E1394 | RS-232 serial, bridged to the network with a serial device server | Driver built |
| Wondfo Finecare FIA Meter Plus (FS-113) | Confirmed at setup | RS-232 serial | Driver built |
| Wondfo Finecare FS-114 | HL7 v2 | over the site network (TCP/IP) | Driver built |
Molecular
| Analyser | Protocol | Physical link | Catenix |
|---|---|---|---|
| Abbott ID NOW | HL7 v2; POCT1-A2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Cepheid GeneXpert | HL7 v2; ASTM E1381/E1394 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| QIAGEN QIAstat-Dx | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Roche cobas Liat | HL7 v2; POCT1-A2 | over the site network (TCP/IP) | Driver built |
| SD Biosensor STANDARD M10 | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| bioMérieux BioFire FilmArray | HL7 v2 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| bioMérieux SPOTFIRE | POCT1-A2 | over the site network (TCP/IP) | Driver built |
Urinalysis
| Analyser | Protocol | Physical link | Catenix |
|---|---|---|---|
| ACON Mission U120 Urine Analyser | Confirmed at setup | RS-232 serial | Connection guide (standard protocol) |
| Analyticon Urilyzer 100 Pro | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| ARKRAY AUTION Eleven | ASTM E1381/E1394; Manufacturer's own format | RS-232 serial, bridged to the network with a serial device server | Connection guide (standard protocol) |
| Arkray AUTION Micro | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Macherey-Nagel Uryxxon Relax | Confirmed at setup | Confirmed at setup | Driver built |
| McKesson Consult urine analyzer | Confirmed at setup | Confirmed at setup | Driver built |
| Roche Urisys 1100 | ASTM E1381/E1394 | RS-232 serial, bridged to the network with a serial device server | Connection guide (standard protocol) |
| Roche cobas u 411 | ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Siemens CLINITEK Advantus | ASTM E1381/E1394 | over the site network (TCP/IP) | Connection guide (standard protocol) |
| Siemens CLINITEK Novus | HL7 v2; ASTM E1381/E1394 | RS-232 serial | Connection guide (standard protocol) |
| Siemens CLINITEK Status+ | POCT1-A2 | over the site network (TCP/IP) | Driver built |
Clinical record systems
| System | Region | How results arrive |
|---|---|---|
| CareFlow EPR | UK | Standard HL7 v2 / FHIR result interface |
| Cliniko | UK, International | Connector built |
| Clinisys WinPath Enterprise | UK | Standard HL7 v2 / FHIR result interface |
| Decoda Health | US | Connector built |
| EMIS Web | UK | Standard HL7 v2 / FHIR result interface |
| Epic | UK, US | Built to the vendor's published specification |
| Experity | US | Standard HL7 v2 / FHIR result interface |
| Greenway Health | US | Standard HL7 v2 / FHIR result interface |
| Healthie | US | Connector built |
| MEDENT | US | Standard HL7 v2 / FHIR result interface |
| MEDITECH Expanse | UK | Standard HL7 v2 / FHIR result interface |
| ModMed | US | Standard HL7 v2 / FHIR result interface |
| NextGen | US | Standard HL7 v2 / FHIR result interface |
| Office Practicum | US | Standard HL7 v2 / FHIR result interface |
| Oracle Health Millennium | UK, US | Standard HL7 v2 / FHIR result interface |
| PCC | US | Standard HL7 v2 / FHIR result interface |
| Point and Click | US | Standard HL7 v2 / FHIR result interface |
| Practice Fusion | US | Standard HL7 v2 / FHIR result interface |
| Semble (formerly Heydoc) | UK | Connector built |
| SystmOne | UK | Standard HL7 v2 / FHIR result interface |
| Tebra | US | Standard HL7 v2 / FHIR result interface |
| Veradigm EHR | US | Standard HL7 v2 / FHIR result interface |
| Vision | UK | Standard HL7 v2 / FHIR result interface |
| athenahealth | US | Built to the vendor's published specification |
| eClinicalWorks | US | Built to the vendor's published specification |