Connecting a Roche cobas b 101 to your EHR or LIS

Quick answer

Yes. Roche describes bidirectional connectivity for the cobas b 101 using the POCT1-A standard, and the operator's manual sends network traffic through its Base Unit Hub port. Catenix connects on that side, receives HbA1c, lipid and CRP results, and sends them to your EHR or LIS.

The cobas b 101 is a small benchtop analyser that runs single-use discs for HbA1c, a lipid panel and CRP. Roche Canada pitches it at pharmacies, physician offices, outpatient clinics and satellite labs, the places where a diabetes or cardiovascular review happens in one visit.

Sample sizes are small: 2 microlitres for HbA1c, 19 for lipids and 12 for CRP, applied straight to the disc. A single HbA1c or lipid test takes about six minutes, CRP under four, and an HbA1c plus lipid pair under fifteen. The lipid disc reports total cholesterol, triglycerides and HDL, with LDL calculated.

WhatHow Catenix handles it
Disc menuHbA1c, lipid panel, CRP
Sample volume2 uL HbA1c, 19 uL lipids, 12 uL CRP
Time per discAbout 6 min (HbA1c, lipids); under 4 min (CRP)
Network portBUH (Base Unit Hub) terminal
Other portsUSB for a PC, USB for stick or printer, barcode scanner terminal
Memory5,000 patient and 500 control results

Choosing BUH over USB

Under Setup, Options, Computer, the b 101 offers three choices: Disabled, USB for one directly attached PC, or BUH to reach a network or data manager through a Base Unit Hub. For an EHR link, BUH is the setting to pick, and the manual points to POCT1-A for that connection. The manual also warns against plugging other devices into the BUH socket or using a multiport USB hub.

Results exported to a USB stick still work for archiving, but they do not reach the patient record on their own.

Per-operator QC on a shared analyser

The b 101 can require each operator to have run QC within a set period, weekly, monthly or every few months, before they can test patients. It also has an instrument-wide QC lockout, an optical check lockout, and a limited number of STAT tests for urgent cases.

Control results that reach Catenix are filed apart from patient results, graded with Westgard rules and charted on Levey-Jennings plots, and a failed control holds patient results for that test. Operator ID stays attached to each result.

Mapping the panel

HbA1c, each lipid fraction and CRP are mapped to their EHR codes during setup. Results without a matching patient wait for review in Catenix, and every message is stored with its result.

Roche and cobas b 101 are trademarks of their owners. Catenix is an independent connectivity provider, not affiliated with or endorsed by Roche.

Questions people ask

Which port on the cobas b 101 connects to a network?

The BUH terminal, through a Base Unit Hub. The USB 1 port is for a single directly attached computer.

Is the cobas b 101 bidirectional?

Roche describes it as bidirectional over POCT1-A. What is exchanged beyond results is confirmed during setup.

Can the cobas b 101 lock out an operator who has not run QC?

Yes. Its Operator Lockout setting requires each operator to run QC within a chosen interval before patient testing.

Related: Abbott Afinion 2 connection · Siemens DCA Vantage connection · QC lockout and operator competency · Device library

Product and company names are trademarks of their owners and are used for identification only. Catenix moves, stores and displays results with traceability; it does not interpret results or give clinical advice. Reviewed November 2026.